FT 1030 Deep-Level Transient Spectroscopy System
|
Products
- Hall
- DLTS
- Customized
Company
- Profile
- Contact
- Partners
- Imprint
- Privacy Policy
Support
- Information
- Downloads
|
Hardware
| Pulse Generator |
| Voltage Range |
±20.4 V (±102 V opt.) |
| Voltage Resolution |
0.625 mV |
| Max. Current |
> ±15 mA |
| Pulse Width |
1 µs - 1000 s |
| Fast Pulse Option available |
| |
| Capacitance Meter |
| HF signal |
100 mV @ 1 MHz (20 mV optional) |
| Ranges [pF] |
3, 30, 300, 3000 (auto or manual) |
| Capacitance compensation |
0.1 - 3000 pF (auto or manual) |
| Sensitivity |
0.01 fF |
| |
| Voltage measurement |
| Range |
±10 V |
| Sensitivity |
< 1 µV |
| Input resistance |
106 Ohm (opt. >1015 Ohm) |
| Misalignment voltage compensation provided |
| |
| Current measurement |
| Ranges |
5, from ±1 µA upto ±10 mA |
| Sensitivity |
< 1 pA |
| Leakage current suppression provided |
| |
| Transient recorder |
| Sample rates |
2 µs upto 2000 s |
| Number of samples |
16 - 16384 (opt. 64k) |
| Adjustable anti-aliasing filter |
| |
| Standard Cryostat |
| Temperature Range |
77 K - 450 K |
| |
| Typical performance (Schottky Diode, Reverse Bias Capacitance 100pF@0V) |
| Sensitivity |
10-7 < NT / (ND - NA) < 10-5 |
| Energy Accurancy |
HT +/- 3% |
| Energy Resolution |
10 meV |
| Emission Rate |
10-3/s < en < 104/s |
|
Options
- Constant Capacitance
- Optical Excitation
- Fast Pulse Interface
- ±100 V Option
- Multi Sample Interface
|